Contact an Account ManagerClean-up of PAHs from Soil
Summary
Type: SPE
Application #: 303860
Catalog#: 730135 - SPE Cartridges, Chromabond, CN/Silica, 6ml, PP, 500/1000mg 30/pk
Detection: After the filtration with CHROMABOND PTFE syringe filter (MN-Cat. No. 729008) the sample was determinated by HPLC according Application No.: 101710
Conditions: Column conditioning:
1 column volume dichloromethane and 1 column volume cyclohexane
Sample application:
Apply the prepared sample on the column
Washing:
2 x 2 ml cyclohexane
Elution:
5 x
Authors: Dreyer Dr
Source: Gerling Institute for Pollution Research 1989
Details